Remote control commands
R&S
®
RTP
1589User Manual 1337.9952.02 ─ 12
BUS<m>:SPI:MOSI:SOURce <MOSISource>
Sets the input channel of the MOSI line.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<MOSISource> NONE | C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 |
M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | D0 | D1 | D2 | D3 | D4 |
D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15
Digital and analog channels cannot be used at the same time for
data, clock and slave select lines.
See Chapter 22.4.2, "Waveform parameter", on page 1145
*RST: C1W1
Usage: Asynchronous command
BUS<m>:SPI:MOSI:POLarity <MOSIPolarity>
Selects whether transmitted data is high active (high = 1) or low active (low = 1).
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<MOSIPolarity> ACTLow | ACTHigh
*RST: ACTHigh
Usage: Asynchronous command
BUS<m>:SPI:SCLK:THReshold <SCLKThreshold>
BUS<m>:SPI:MISO:THReshold <MISOThreshold>
BUS<m>:SPI:MOSI:THReshold <MOSIThreshold>
BUS<m>:SPI:SSELect:THReshold <SSThreshold>
Set user-defined threshold values for the clock, MISO, MOSI and slave select lines.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<SSThreshold> User-defined value
Range: -12 to 12
Increment: 0.1
*RST: 0
Default unit: V
Protocols