Remote control commands
R&S
®
RTP
1795User Manual 1337.9952.02 ─ 12
TRIGger<m>:TBTO:ERRor:CRC <ErrorCRC>
Enables triggering on cyclic redundancy check (CRC) error.
Suffix:
<m>
.
1..3
Parameters:
<ErrorCRC> ON | OFF
*RST: ON
TRIGger<m>:TBTO:ERRor:FEC <ErrorFEC>
Enables triggering, if more than 22 RS-FEC errors are detected.
Suffix:
<m>
.
1..3
Parameters:
<ErrorFEC> ON | OFF
*RST: ON
TRIGger<m>:TBTO:ERRor:OOR <ErrorOOR>
Enables triggering if values are out of range (OOR). For example, a preamble error, or
invalid symbols in control blocks.
Suffix:
<m>
.
1..3
Parameters:
<ErrorOOR> ON | OFF
*RST: ON
TRIGger<m>:TBTO:ERRor:ZERO <ErrorZERO>
Enables triggering, if a zero ternary symbol pair (0-0) is detected.
Suffix:
<m>
.
1..3
Parameters:
<ErrorZERO> ON | OFF
*RST: ON
TRIGger<m>:TBTO:LENGth:CONDition <TypeOperator>
Sets the condition for the length / type. You can define an exact value or a value range.