Remote control commands
R&S
®
RTP
1848User Manual 1337.9952.02 ─ 12
Parameters:
<ThresholdHSDP3> Range: -1 to 1
Increment: 0.01
*RST: 0.2
Default unit: V
BUS<m>:DPHY:SMLanes <Active>
Enables the display of the results in multiple lanes.
Suffix:
<m>
.
1..4
Parameters:
<Active> ON | OFF
*RST: ON
22.17.14.2 D-PHY trigger
The trigger suffix <m> is always 1 and can be omitted. It selects the trigger event: Only
the A-trigger is available for triggering on serial buses.
To trigger on a serial bus, make sure that:
●
TRIGger<m>:SOURce[:SELect] is set to SBUS.
●
The sources of the serial bus are channel signals: use BUS<m>:...:SOURce
commands.
●
Decoding is enabled: BUS<m>[:STATe] is set to ON.
TRIGger<m>:DPHY:TYPE............................................................................................ 1849
TRIGger<m>:DPHY:DATA:CONDition............................................................................ 1849
TRIGger<m>:DPHY:DATA:MAX.....................................................................................1850
TRIGger<m>:DPHY:DATA:MIN......................................................................................1850
TRIGger<m>:DPHY:DIDX:CONDition.............................................................................1850
TRIGger<m>:DPHY:DIDX:MAX..................................................................................... 1850
TRIGger<m>:DPHY:DIDX:MIN...................................................................................... 1851
TRIGger<m>:DPHY:DTYPe:CONDition..........................................................................1851
TRIGger<m>:DPHY:DTYPe:MAX.................................................................................. 1851
TRIGger<m>:DPHY:DTYPe:MIN................................................................................... 1851
TRIGger<m>:DPHY:ESCMode:CONDition......................................................................1852
TRIGger<m>:DPHY:ESCMode:MAX..............................................................................1852
TRIGger<m>:DPHY:ESCMode:MIN............................................................................... 1852
TRIGger<m>:DPHY:ESDTa:CONDition.......................................................................... 1852
TRIGger<m>:DPHY:ESDTa:MAX...................................................................................1853
TRIGger<m>:DPHY:ESDTa:MIN....................................................................................1853
TRIGger<m>:DPHY:ESINdex:CONDition........................................................................1853
TRIGger<m>:DPHY:ESINdex:MAX................................................................................1854
TRIGger<m>:DPHY:ESINdex:MIN................................................................................. 1854
TRIGger<m>:DPHY:HSVC............................................................................................1854
Protocols