Acquisition and waveform setup
R&S
®
RTP
186User Manual 1337.9952.02 ─ 12
Figure 5-15: Probe setup for current probes R&S
RT-ZC20B
For all current probes, attenuation or gain is shown in the "Setup" tab, and you can set
the offset to mean:
●
"Probe unit, Auto attenuation" on page 168
●
"Offset to mean" on page 184
Current probes R&S RT-ZCxxB are adjusted by the following functions:
DeGauss..................................................................................................................... 186
Detect AutoZero..........................................................................................................187
Zero adjust.................................................................................................................. 187
Save to probe..............................................................................................................187
DeGauss
Demagnetizes the core if it has been magnetized by switching the power on and off, or
by an excessive input. Always carry out demagnetizing before measurement.
The demagnetizing process takes about one second. During demagnetizing, a demag-
netizing waveform is displayed.
Demagnetizing is done automatically when R&S RT-ZCxxB is connected to the oscillo-
scope, or when "Detect AutoZero" is performed.
Remote command:
PROBe<m>:SETup:DEGauss on page 1226