Remote control commands
R&S
®
RTP
1924User Manual 1337.9952.02 ─ 12
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less than or equal, Greater than,
Greater than or equal. These conditions require one endpoint
value to be set using TRIGger<m>:MPHY:SSIC:FRAMe<n>:
FLD<o>:DMIN.
INRange | OORange
In range, out of range. These conditions require a range of end-
point values to be set using TRIGger<m>:MPHY:SSIC:
FRAMe<n>:FLD<o>:DMIN and TRIGger<m>:MPHY:SSIC:
FRAMe<n>:FLD<o>:DMAX.
*RST: EQUal
TRIGger<m>:MPHY:SSIC:IMAX
<Frame>, <Field>, <Data>
TRIGger<m>:MPHY:SSIC:FRAMe<n>:FLD<o>:IMAX <IndexMax>
Sets the end value of an index range if TRIGger<m>:MPHY:SSIC:FRAMe<n>:
FLD<o>:IOPerator is set to INRange or RANGe.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
<n> *
Specifies the field number within the frame.
<o> *
Parameters:
<IndexMax> Specifies the end value for the index range within the field.
The index range, increment and *RST values depend on the
field type.
TRIGger<m>:MPHY:SSIC:IMIN <Frame>, <Field>, <Data>
TRIGger<m>:MPHY:SSIC:FRAMe<n>:FLD<o>:IMIN <IndexMin>
Specifies the index, or sets the start value of an index range.
Suffix:
<m>
.
1..3
<n> *
Specifies the field number within the frame.
<o> *
Parameters:
<IndexMin> Specifies the index value or index start value within the field.
The index range, increment and *RST values depend on the
field type.
Range: 1 to 65534
Increment: 1
*RST: 1