Remote control commands
R&S
®
RTP
1933User Manual 1337.9952.02 ─ 12
NRZU
NRZ (non-return-to-zero), unclocked
*RST: MANC
BUS<m>:CMSB:MANChester:DATA <SourceMANData>
Selects the source channel for the data signal. For triggering on a serial bus, analog
channels "C1""C4" are required. Otherwise, if no serial bus trigger has been selected,
permitted source selections include the mathematical channels and the reference
channels .
Suffix:
<m>
.
1..4
Parameters:
<SourceMANData> C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 | M5 | M6 |
M7 | M8 | R1 | R2 | R3 | R4
*RST: C1W1
BUS<m>:CMSB:MANChester:POLarity <PolarityData>
Selects the polarity of the custom serial bus data signal in Manchester coding stand-
ards.
Suffix:
<m>
.
1..4
Parameters:
<PolarityData> NORMal | INVert
NORMal
Manchester or Manchester II polarity remains unchanged
INVert
Manchester polarity is inverted and becomes Manchester II
polarity, Manchester II polarity is inverted and becomes Man-
chester polarity
*RST: NORMal
BUS<m>:CMSB:MANChester:THReshold:HIGH <ThresholdUpper>
Sets the upper threshold for data signal digitization in Manchester coding standards.
Suffix:
<m>
.
1..4
Parameters:
<ThresholdUpper> Range: -25 to 25
Increment: 0.1
*RST: 5
Default unit: V