Loading...
Remote control commands
R&S
®
RTP
1957User Manual 1337.9952.02 ─ 12
<n> *
<o> *
Parameters:
<IndexOperator> EQUal | INRange | RANGe
*RST: INRange
22.17.16.4 Decode results
In all BUS<m>:CSMB:RESult<n> commands, the suffix <m> selects the serial bus, the
suffix <n> selects the result number in the decode table, and the suffix <o> selects the
cell number.
The following is an example for a set of query commands for bus 1 and result 1 is
shown in the following, together with examples for outcomes of these queries:
:BUS1:CMSB:RCOunt? !5
:BUS1:CMSB:RESult1:STATe? !OK
:BUS1:CMSB:RESult1:STARt? ! -0.0024964177
:BUS1:CMSB:RESult1:STOP? ! -0.0024030384
:BUS1:CMSB:RESult1:TYPE? !ff
:BUS1:CMSB:RESult1:CONE? !0b11111111
:BUS1:CMSB:RESult1:CTWO? !0xAA
:BUS1:CMSB:RESult1:CTHR? !0xF590
:BUS1:CMSB:RESult1:CCOunt? !5
:BUS1:CMSB:RESult1:CELL1:STAT? !OK
:BUS1:CMSB:RESult1:CELL1:NAME? !Start Delim
:BUS1:CMSB:RESult1:CELL1:VALue? !101010101HL10HL0
:BUS1:CMSB:RESult1:CELL2:STAT? !OK
:BUS1:CMSB:RESult1:CELL2:NAME? !OP-FF
:BUS1:CMSB:RESult1:CELL2:VALue? !0b11111111
:BUS1:CMSB:RESult1:CELL3:STAT? !OK
:BUS1:CMSB:RESult1:CELL3:NAME? !data
:BUS1:CMSB:RESult1:CELL3:VALue? !0xAA
:BUS1:CMSB:RESult1:CELL4:STAT? !OK
:BUS1:CMSB:RESult1:CELL4:NAME? !CRC
:BUS1:CMSB:RESult1:CELL4:VALue? !0xF590
Protocols