EasyManua.ls Logo

Rohde & Schwarz RTP Series - Page 1977

Rohde & Schwarz RTP Series
2403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTP
1977User Manual 1337.9952.02 ─ 12
BUS<m>:EBTB:DPLus:SOURce <SourceDplus>
Selects the D+ source of the provided single-ended signal, if BUS<m>:EBTB:TYPE
SINGle is set.
Suffix:
<m>
.
1..4
Parameters:
<SourceDplus> NONE | C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 |
M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4
*RST: C1W1
Usage: Asynchronous command
BUS<m>:EBTB:DPLus:THReshold
<ThresholdDplus>
Sets the high threshold (D +) of the signal, if BUS<m>:EBTB:TYPE SINGle.
Suffix:
<m>
.
1..4
Parameters:
<ThresholdDplus> Range: -10 to 10
Increment: 1E-3
*RST: 0
Default unit: V
BUS<m>:EBTB:FCSY <CustSync>
Sets a pattern value, if BUS<m>:EBTB:SYNC CUS.
Suffix:
<m>
.
1..4
Parameters:
<CustSync>
BUS<m>:EBTB:SCOunt?
Returns the symbol count for the selected serial bus, i.e. the number of symbols in the
present acquisition.
Suffix:
<m>
.
1..4
Return values:
<Count> Range: 0 to 100000
Increment: 1
*RST: 0
Usage: Query only
Protocols

Table of Contents

Other manuals for Rohde & Schwarz RTP Series

Related product manuals