Remote control commands
R&S
®
RTP
1990User Manual 1337.9952.02 ─ 12
BUS<m>:MDIO:CLOCk:SOURce <SourceClock>
Selects the source for the clock line (management data clock, MDC). Permitted selec-
tions are the analog channels "C1""C4" and the digital channels "D0""D15".
Suffix:
<m>
.
1..4
Parameters:
<SourceClock> C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 | M5 | M6 |
M7 | M8 | R1 | R2 | R3 | R4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 |
D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15
Digital and analog channels cannot be used at the same time for
data and clock lines. For triggering on a serial bus, analog or
digital input channels are required.
See Chapter 22.4.2, "Waveform parameter", on page 1145
*RST: C2W1
BUS<m>:MDIO:DATA:SOURce
<SourceData>
Selects the source for the data signal. Permitted selections are the analog channels
"C1""C4" and the digital channels "D0""D15", but not the same as for "Clock".
Suffix:
<m>
.
1..4
Parameters:
<SourceData> C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 | M5 | M6 |
M7 | M8 | R1 | R2 | R3 | R4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 |
D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15
Digital and analog channels cannot be used at the same time for
data and clock lines. For triggering on a serial bus, analog or
digital input channels are required.
See Chapter 22.4.2, "Waveform parameter", on page 1145
*RST: C1W1
BUS<m>:MDIO:CLOCk:THReshold:HIGH <ThresClkHigh>
Defines the upper threshold level for the clock signal.
Suffix:
<m>
.
1..4
Parameters:
<ThresClkHigh> Range: -5 to 5
Increment: 0.1
*RST: 2
Default unit: V