Remote control commands
R&S
®
RTP
2069User Manual 1337.9952.02 ─ 12
BUS<m>:USBThree:RESult:FRAMe<n>:STOP?
Returns the stop time of the frame with the selected frame number (for the selected
serial bus).
Suffix:
<m>
.
1..4
<n> *
Selects the frame number.
Return values:
<Stop> Range: -100E+24 to 100E+24
Increment: 100E-12
Default unit: s
Usage: Query only
BUS<m>:USBThree:RESult:FRAMe<n>:FLD<o>:TYPE?
Returns the type of the field with the selected field number within the frame with the
selected frame number (for the selected serial bus).
Suffix:
<m>
.
1..4
<n> *
Selects the frame number.
<o> *
Selects the field number.
Return values:
<FieldType>
Usage: Query only
BUS<m>:USBThree:RESult:FRAMe<n>:FLD<o>:STATus?
Returns the status of the field with the selected field number within the frame with the
selected frame number (for the selected serial bus).
Suffix:
<m>
.
1..4
<n> *
Selects the frame number.
<o> *
Selects the field number.
Protocols