Remote control commands
R&S
®
RTP
2094User Manual 1337.9952.02 ─ 12
FBFB
4b5b error
*RST: OK
Usage: Query only
BUS<m>:USBPd:RESult:FRAMe<n>:STOP?
Returns the stop time of the frame with the selected frame number (for the selected
serial bus).
Suffix:
<m>
.
1..4
<n> *
Return values:
<Stop> Range: -100E+24 to 100E+24
Increment: 100E-12
*RST: 0
Default unit: s
Usage: Query only
BUS<m>:USBPd:RESult:FRAMe<n>:TYPE?
Returns the frame type for the selected serial bus and frame number.
Suffix:
<m>
.
1..4
<n> *
Return values:
<FrameType> STARt | RESet | CTRL | DATA | BIST | RQST | SINK | SRC |
VEND | BATT | ALRT | XMSG | XSRC | XSTA | XGBC | XGBS |
XBAC | XGMI | XMFI | XSRQ | XSRS | XFRQ | XFRS | TEST |
LOWP | UNKNown
For a description of the frame types, see "Frame packet types"
on page 849.
*RST: UNKNown
Usage: Query only
22.17.21.4 Search settings
The trigger suffix <m> is always 1 and can be omitted. It selects the trigger event: Only
the A-trigger is available for triggering on serial buses.
To trigger on a serial bus, make sure that:
TRIGger<m>:SOURce[:SELect] is set to SBUS.
Protocols