Triggers
R&S
®
RTP
221User Manual 1337.9952.02 ─ 12
Polarity
Indicates the polarity of a pulse, that is the direction of the first pulse slope.
"Positive"
Selects positive going pulses.
"Negative"
Selects negative going pulses.
"Either"
Selects both positive and negative going pulses.
Remote command:
TRIGger<m>:GLITch:POLarity on page 1259
TRIGger<m>:RUNT:POLarity on page 1262
Upper level
Sets the upper voltage threshold.
Remote command:
TRIGger<m>:LEVel<n>:RUNT:UPPer on page 1262
Lower level
Sets the lower voltage threshold.
Remote command:
TRIGger<m>:LEVel<n>:RUNT:LOWer on page 1262
Range
Selects how the time limit of the runt pulse is defined:
"Any runt"
Triggers on all runts fulfilling the level condition, without time limita-
tion.
"Longer"
Triggers on runts longer than the given "Runt width".
"Shorter"
Triggers on runts shorter than the given "Runt width".
"Within"
Triggers if the runt length is inside a given time range. The range is
defined by "Runt width" and "±Delta".
"Outside"
Triggers if the runt length is outside a given time range. The range
definition is the same as for "Within" range.
Remote command:
TRIGger<m>:RUNT:RANGe on page 1262
Trigger types