Remote control commands
R&S
®
RTP
2265User Manual 1337.9952.02 ─ 12
Suffix:
<m>
.
1..2
Return values:
<DatarateValue> Range: 1000 to 10E+9
Increment: 10
*RST: 1E+9
Default unit: bps
Usage: Query only
ADVJitter<m>:SIGNal:SAMTime <SamplingTime>
Sets a sampling time for the explicit clock signal, an offset for the clock edge in relation
to the bit start. The clock edge sets the beginning of the unit interval. The sampling
time is a number between 0 and 1. Value 0 sets the clock edge to the beginning of the
bit period; value 0.5 sets the clock edge to the middle of the bit period.
Suffix:
<m>
.
1..2
Parameters:
<SamplingTime> Range: 0 to 1
Increment: 0.01
*RST: 0.5
Default unit: UI
ADVJitter<m>:SIGNal:GATE:COUPling <ModeType>
Sets if the gate is coupled to the zoom or cursor.
Suffix:
<m>
.
1..2
Parameters:
<ModeType> NONE | CURSor | ZOOM
*RST: NONE
ADVJitter<m>:SIGNal:GATE:CURSor <CursorType>
Sets the cursor the gate is coupled to, if ADVJitter<m>:SIGNal:GATE:COUPling
is set to CURSor.
Suffix:
<m>
.
1..2
Parameters:
<CursorType> CURSOR1 | CURSor1 | CURSOR2 | CURSor2 | CURSOR3 |
CURSor3 | CURSOR4 | CURSor4
*RST: CURSOR1
Jitter analysis and clock data recovery