Triggers
R&S
®
RTP
228User Manual 1337.9952.02 ─ 12
Clock source
Selects the input channel of the clock signal.
Remote command:
TRIGger<m>:SETHold:CSOurce[:VALue] on page 1272
Clock edge
Sets the edge of the clock signal to define the time reference point for the setup and
hold time:
"Positive"
Rising edge, a positive voltage change.
"Negative"
Falling edge, a negative voltage change.
"Both"
Both the rising and the falling edge.
Remote command:
TRIGger<m>:SETHold:CSOurce:EDGE on page 1271
Clock level
Sets the voltage level for the clock signal. Both "Clock level" and "Clock edge" define
the starting point for calculation of the setup and hold time.
Remote command:
TRIGger<m>:SETHold:CSOurce:LEVel on page 1271
Data level
Sets the voltage level for the data signal. At this level, the setup and hold time is mea-
sured.
Remote command:
TRIGger<m>:LEVel<n>[:VALue] on page 1255