Remote control commands
R&S
®
RTP
2298User Manual 1337.9952.02 ─ 12
<n> 1..n
Index of the component in the measurement setup. The DUT
has index 1, the probe has index n.
Return values:
<TotalPorts> ONE | TWO | THRee | FOUR
*RST: TWO
Usage: Query only
Asynchronous command
22.23.3.2 DUT settings
DEEMbedding<m>:COMPonent<n>:MODE....................................................................2298
DEEMbedding<m>:COMPonent<n>:LOAD:REMProbeload..............................................2298
DEEMbedding<m>:COMPonent<n>:LOAD:IMPedance?..................................................2299
DEEMbedding<m>:COMPonent<n>:MODE <Mode>
Sets the probing mode.
Suffix:
<m>
.
1..4
Selects the channel that is configured for deembedding.
<n> 1, only for available DUT
Parameters:
<Mode> PROBing | TERMinating
PROBing
For measurements with a high-ohmic probe and measure on an
existing line in parallel to the load.
TERMinating
For measurements if the measuring equipment is the load of the
line.
*RST: PROBing
Usage: Asynchronous command
DEEMbedding<m>:COMPonent<n>:LOAD:REMProbeload <Value>
Removes the loading of the probe if DEEMbedding<m>:COMPonent<n>:MODE is set
to PROBing.
If the probe loading is removed, you see the signal that would be at the measurement
point if the probe's input impedance is ideal (infinite impedance).
Without removing the loading, you see the real signal at the measurement point,
including the probe loading.
Deembedding (Option R&S
RTP-K121/-K122)