Triggers
R&S
®
RTP
248User Manual 1337.9952.02 ─ 12
The reset condition R can be a timeout or a trigger condition, or a combination of both.
Sequence
Selects the type of the sequence.
"A B R"
Triggers if all conditions of A- and B-events, as well as additional
delay and reset timeout or R-event (reset) conditions are fulfilled.
"A Zone"
Triggers if the conditions of the A-event and the zone trigger are fulfil-
led. A holdoff condition also can be set.
"A B R
Zone"
Triggers if all conditions of A- and B-events, reset timeout or R-event
(reset) conditions are fulfilled for the defined zone.
"A OR B"
Triggers if the conditions of A-event or the conditions of the B-event
are fulfilled.
"A OR B
Zone"
Triggers if the conditions of A-event or the conditions of the B-event
are fulfilled for the defined zone. A holdoff condition also can be set.
Remote command:
TRIGger<m>:SEQuence:TYPE on page 1310
After 1st event wait
Sets the time that the instrument waits after an A-trigger until it recognizes B-triggers.
Remote command:
TRIGger<m>:SEQuence:DELay on page 1311
2nd event count
Sets the number of B-trigger conditions to be fulfilled after an A-trigger. The last B-trig-
ger causes the trigger event.
The waiting time for B-triggers can be restricted with a reset condition: timeout or reset
event.
Only available in "A B R" sequences.
Remote command:
TRIGger<m>:SEQuence:COUNt on page 1311
Sequence