Measurements
R&S
®
RTP
337User Manual 1337.9952.02 ─ 12
Long-term measurements are performed on the measurement that is selected for
analysis and math on the "Result Analysis" tab.
See: Chapter 8.2.11, "Result analysis", on page 387
●
Intermediate results
You can display auxiliary result lines and reference levels in the source diagram,
see Chapter 8.2.2.2, "Configuring the results display", on page 338.
Remote commands:
●
MEASurement<m>:ARES? on page 1366
●
MEASurement<m>:ARNames on page 1367
●
MEASurement<m>:RESult[:ACTual]? on page 1367
●
MEASurement<m>:RESult:COUNt? on page 1369
8.2.2.1 Measurement status
The overall status of measurement results is indicated by various icons. In general, a
question mark before the result value indicates that the measurement result might not
be correct due to insufficient amplitude level. Check your amplitude and reference level
settings. The icon colors indicate the state of the limit and margin checks.
Icon Description
No icon, no result
value ("---")
The instrument cannot measure the required value, for example, if the acquisition
does not contain at least one complete period for frequency and cycle measure-
ments. Check and adjust the waveform settings to get results.
The measurement result might not be correct due to insufficient amplitude level.
Check your amplitude and reference level settings.
Limit and margin checks are disabled.
Limit and margin checks passed, measurement results are reliable.
The measurement result might not be correct due to insufficient amplitude level.
Check your amplitude and reference level settings.
Limit and margin checks passed.
The measurement result might not be correct due to insufficient amplitude level.
Check your amplitude and reference level settings.
Margin checks failed.
Margin checks failed.
The measurement result might not be correct due to insufficient amplitude level.
Check your amplitude and reference level settings.
Limit checks failed.
Limit checks failed.
Automatic measurements