Mask testing
R&S
®
RTP
431User Manual 1337.9952.02 ─ 12
Rescale
You can rescale and move mask segments by numerical input of factors and offsets.
The values change the selected mask segment and take effect on "Apply".
Offset X ← Rescale
Moves the mask segment horizontally. The specified offset is added to the x-values of
all points of the selected mask segment.
To take effect, tap "Apply".
Remote command:
MTESt:SEGMent:RESCale:XOFFset on page 1468
Factor X ← Rescale
Stretches or compresses the selected mask segment in horizontal direction. The x-val-
ues of all points of the selected mask segment are multiplied with this factor. Factors
>1 stretch the mask segment, while factors between 0 and 1 compress it. Negative val-
ues are possible and change the algebraic sign.
To take effect, tap "Apply".
Remote command:
MTESt:SEGMent:RESCale:XFACtor on page 1468
Offset Y ← Rescale
Moves the mask segment vertically. The specified offset is added to the y-values of all
points of the selected mask segment.
To take effect, tap "Apply".
Remote command:
MTESt:SEGMent:RESCale:YOFFset on page 1469
Factor Y ← Rescale
Stretches or compresses the selected mask segment in vertical direction. The y-values
of all points of the selected mask segment are multiplied with this factor. Factors >1
stretch the mask segment, while factors between 0 and 1 compress it. Negative values
are possible and change the algebraic sign.
To take effect, tap "Apply".
Remote command:
MTESt:SEGMent:RESCale:YFACtor on page 1468
Apply ← Rescale
Multiplies and adds the given x- and y-factors and offsets to the coordinates of all
points of the selected mask segment.
Remote command:
MTESt:SEGMent:RESCale:RECalculate on page 1467
10.2.2.2 Mask definition: waveform mask
Access: [App Cockpit] > "Analysis" tab > "Define Mask" > "Mask Type" = "Waveform" >
"Setup mask"
Mask test settings