Mask testing
R&S
®
RTP
444User Manual 1337.9952.02 ─ 12
d) To add or delete the selected point, use the buttons below the table.
● "Insert": to add a new point before the selected point.
● "Append": to add a new point at the end of the list.
● "Remove": to delete the selected point from the list.
e) To move the selected point, change the X- and Y-values.
To rescale and move a mask segment
The settings mentioned here are described in detail in Chapter 10.2.2, "Mask defini-
tion", on page 428.
1. Open [App Cockpit] > "Analysis" tab > "Mask" > "Define Mask".
2. At the top, select the mask test for which you want to change the mask definition.
3. Tap "Setup mask" and select the required segment in the "User Mask: Segments"
table.
4. Tap "Configure".
5. To stretch or compress the selected mask segment, enter the "Factor X" for hori-
zontal scaling and the "Factor Y" for vertical scaling. The x-values and y-values of
all points are multiplied with the corresponding factor. Factors >1 stretch the mask
segment, while factors between 0 and 1 compress it. Negative values are possible
and change the algebraic sign.