Data and file management
R&S
®
RTP
501User Manual 1337.9952.02 ─ 12
2. Tap the "Waveform" button.
The "Save - Waveform" dialog opens.
3. Select the waveforms to be saved:
To save one waveform, tap the "Source" icon and select the waveform.
To save data of several channels, enable "Multi channel" and select the chan-
nels.
4. In the "Scope" list, select the part of the waveform record to be saved. Zoom, cur-
sor and gate segments require the same setup for the selected waveform before
saving. For "Manual", enter the "Start" and "Stop" time of the section.
5. Check the filename under "Save to file" and change it if needed. Usually, autonam-
ing is used.
6. Check the file format and the "Export format" settings and change them if needed.
7. Tap "Save" to save the waveform data to the specified file.
Tap "Save As" to save the waveform data to a different file or file type. Select the
file from the file selection dialog box.
To save a waveform using the toolbar icon
1. Add the "Save Waveform" icon to the toolbar, see Chapter 3.3.6.2, "Configuring the
toolbar", on page 74.
2. Set the scope, export format, and other parameters in "Menu" > "Save/Recall" >
"Waveforms".
3. If necessary, adjust the autonaming pattern and the storage path in "Menu" > "Set-
tings" > "Save/Recall" > "Autonaming".
4. To save the waveform data:
a) Tap the "Save Waveform" icon on the toolbar. The "Save Waveform" toolbar
assist opens.
b) Tap the waveform you want to save.
If you tap the diagram background, the data of the focused waveform is saved.
To export waveform data of a running acquisition
1. Select the waveform you want to save and the scope as described in "To save a
waveform or a part of a waveform to a file" on page 500, step 1 to 5.
2. If you want to save only a section of each waveform, set the "Scope".
3. Open the "Logger" tab.
4. Tap "Enable".
5. Enter the number of acquisitions to be acquired and saved in "Acquisition count".