587User Manual 1337.9952.02 ─ 12
Nodes {
Master: Master1, 10 ms, 0.1 ms;
Slaves: Slave_Motor, Slave_Sensor;
}
Signals {
Sensor1:1,0x0,Slave_Sensor, Master1;
Temperature1:12,0,Slave_Sensor, Master1;
}
Frames {
Sensor_Frame: {Error,30;
Temperature, 20;
Status, 0;}
Master_Frame: {Error,15;
Speed, 10;}
}
Schedule_tables {
Sensor_Frame delay 10.000 ms;
Master_Frame delay 15.000 ms;
}
Signal_encoding_types {
Speed {logical_value, 0, "Motor_off";
logical_value, 1, "Speed1";
logical_value, 2, "Speed2";
logical_value, 3, "Speed3";
}
}
Signal_representation {
Speed1_encoding:Speed1
Temperature1_encoding:Temperature1
}
Remote commands:
BUS<m>:LIN:FRAMe<n>:SDATa? on page 1667
BUS<m>:LIN:FRAMe<n>:SDEXport? on page 1667
13.6.2.3 Configuring LIN signals
For configuration, assign the lines to the input channels, and define the active states
and the logical thresholds.
For details on configuration settings, see Chapter 13.6.2.1, "LIN configuration settings",
on page 583.
1. Press the [Protocol] key on the front panel.
LIN (option R&S
RTP-K3)