Protocol analysis
R&S
®
RTP
600User Manual 1337.9952.02 ─ 12
See also: Chapter 13.1.1, "Setup - general settings", on page 512
Source
Sets the source of the selected data line. Waveform 1 of channel signals, math wave-
forms, and reference waveforms can be used.
For triggering on a serial bus, a channel signal is required.
Remote command:
BUS<m>:MILStd:SOURce on page 1685
Polarity
Selects the wire on which the bus signal is measured : "Normal" or "Inverted". The set-
ting affects the digitization of the signal.
Remote command:
BUS<m>:MILStd:POLarity on page 1687
Thresholds
Threshold values are used for digitization of the signal.
Sets the threshold value for digitization of signals for each line. If the signal value on
the line is higher than the threshold, the signal state is high (1 or true for the Boolean
logic). Otherwise, the signal state is considered low (0 or false) if the signal value is
below the threshold.
MIL-1553 (option R&S
RTP-K6)