EasyManua.ls Logo

Rohde & Schwarz RTP Series - Page 748

Rohde & Schwarz RTP Series
2403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Protocol analysis
R&S
®
RTP
748User Manual 1337.9952.02 ─ 12
Remote command:
BUS<m>:CMSB:NRZ:CLCK on page 1935
Data Polarity (Manchester)
Defines the polarity of the custom serial bus data signal in Manchester coding stand-
ards. The available settings are "Normal" or "Inverted".
Remote command:
BUS<m>:CMSB:MANChester:POLarity on page 1933
Data Polarity (NRZ)
Defines the polarity of the custom serial bus data signal in NRZ coding standards. The
available settings are:
"Active high"
The value "1" is represented by a voltage above the threshold.
"Active low"
The value "1" is represented by a voltage below the threshold.
Remote command:
BUS<m>:CMSB:NRZ:POLarity on page 1938
Data Idle Polarity (NRZ Unclocked)
Defines the idle polarity of the custom serial bus data signal (only available for the cod-
ing standard "NRZ Unclocked"). The available settings are:
"Idle low"
The base value of the clock is "0"; after an idle period, the data signal
starts with a low-to-high transition.
"Idle high"
The base value of the clock is "1"; after an idle period, the data signal
starts with a high-to-low transition.
Remote command:
BUS<m>:CMSB:NRZ:IDLPolarity on page 1936
Clock Polarity (NRZ Clocked)
Defines the polarity of the custom serial bus clock signal (only available for the coding
standard "NRZ Clocked"). The available settings are:
"Idle low"
The base value of the clock is "0".
"Idle high"
The base value of the clock is "1".
Remote command:
BUS<m>:CMSB:NRZ:CPOLarity on page 1936
Clock Phase (Manchester)
Defines the phase of the custom serial bus clock signal for the Manchester coding
standards. The available settings are:
"Auto Detect"
Lets the decoder automatically select the method ("First Edge" or
"Second Edge") for detecting the clock phase.
"First Edge"
At "Idle" = "low": data are captured on the clock's rising edge
(low-to-high transition) and propagated on a falling edge.
At "Idle" = "high": data are captured on the clock's falling edge
(high-to-low transition) and propagated on a rising edge.
Custom: Manchester / NRZ (option R&S RTP-K50)M

Table of Contents

Other manuals for Rohde & Schwarz RTP Series

Related product manuals