Protocol analysis
R&S
®
RTP
767User Manual 1337.9952.02 ─ 12
You can define individual checking parameters for the fields listed in the "Trigger type
dependent settings".
To specify these parameters, select a field from this list and define the data and/or
index operators and values, or the bit state.
The trigger instant is the last criterion that is fulfilled.
Remote command:
TRIGger<m>:CMSB:ADVanced:FRAMe<n>:ENABle on page 1949
Field name / Condition / Enable ← Advanced
This table lists the field numbers and names in the selected frame together with a sum-
mary of the user settings of checking conditions for each field and a checkbox to
enable the checking.
Select a field in the table to specify the checking conditions for this field in the "Data",
"BitState" and/or "Index" dialog (whichever applies). The condition is only applied, and
the "Condition" is only shown in the table, if "Enable" is checked.
Remote command:
TRIGger<m>:CMSB:ADVanced:FRAMe<n>:FLD<o>:ENABle on page 1949
TRIGger<m>:CMSB:ADVanced:FIENable on page 1949
BitState ← Advanced
Defines the bit state to be checked for the selected field. Permitted bit states are "1",
"0" or "X" (do not care).
Remote command:
TRIGger<m>:CMSB:ADVanced:FRAMe<n>:FLD<o>:BIT on page 1950
TRIGger<m>:CMSB:ADVanced:BIT on page 1950
Data ← Advanced
Defines for the selected field, how a data check is executed.
"Condition"
Defining specific data or a data range requires to set the operator to
one of the following conditions: equal, not equal, less than, less than
or equal, greater than, greater than or equal, in range, out of range.
"Min"
Specifies data or sets the start value of a data range.
"Max"
Sets the end value of a data range, if "Condition" is set to "In range"
or "Out of range".
Remote command:
TRIGger<m>:CMSB:ADVanced:DMAX on page 1950
TRIGger<m>:CMSB:ADVanced:FRAMe<n>:FLD<o>:DMAX on page 1950
TRIGger<m>:CMSB:ADVanced:DMIN on page 1951
TRIGger<m>:CMSB:ADVanced:FRAMe<n>:FLD<o>:DMIN on page 1951
TRIGger<m>:CMSB:ADVanced:DOPerator on page 1951
TRIGger<m>:CMSB:ADVanced:FRAMe<n>:FLD<o>:DOPerator on page 1951
Index ← Advanced
Defines for the selected field, how an index check is executed.
"Condition"
Defining a specific index or an index range requires to set the opera-
tor to one of the following conditions: equal, in range.
Custom: Manchester / NRZ (option R&S RTP-K50)M