EasyManua.ls Logo

Rohde & Schwarz RTP Series - Page 884

Rohde & Schwarz RTP Series
2403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Protocol analysis
R&S
®
RTP
884User Manual 1337.9952.02 ─ 12
Make sure that:
In the [Protocol] > "Setup" tab:
The correct "Protocol" is selected.
The data sources of the serial bus are channel signals.
In the "Trigger" > "Setup" tab:
The "Trigger on" = "Single event"
The trigger "Source" = "Serial bus".
The correct "Serial bus" is selected.
Trigger level
Defines the trigger level for PCIe 3.
Remote command:
TRIGger<m>:PCIE:TRGLevel on page 2131
Trigger type PCI express
Selects the trigger type for the PCI express analysis.
"TLP (Transac-
tion layer)"
Sets the trigger on a defined type of the transaction layer packet.
"DLLP (Data
link layer)"
Sets the trigger on a defined type of the data link layer packet.
"Ordered Sets"
Sets the trigger on a defined ordered set.
"Error Condi-
tion"
Sets the trigger on a selection of error conditions.
Remote command:
TRIGger<m>:PCIE:TYPE on page 2140
TLP (Transaction layer)
Specifies the trigger conditions for the transaction layer type trigger.
PCIe (option R&S
RTP-K72/K73)

Table of Contents

Other manuals for Rohde & Schwarz RTP Series

Related product manuals