EasyManua.ls Logo

Rohde & Schwarz RTP044 - Page 1027

Rohde & Schwarz RTP044
2403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Jitter analysis and clock data recovery
R&S
®
RTP
1027User Manual 1337.9952.02 ─ 12
Serial
standard
Nom. bit
rate
Bandwidth Rel. bandwidth Order Damping
Thunder-
bolt 10G
10.3125
Gbps
1667 6.19 MHz 1st -
XAUI 3.75 Gbps 1667 2.25 MHz 1st -
Remote command:
ADVJitter<m>:SIGNal:SERStandard on page 2267
Signal Type
Select, if the analyzed signal is a data or clock signal.
Remote command:
ADVJitter<m>:SIGNal:TYPE on page 2262
Quick start analysis
Starts a quick advanced jitter analysis.
Before starting the quick start analysis, set the Analyzed source and the Signal Type .
The quick start analysis performs the following:
A bit rate estimation is performed on the applied signal.
A software CDR of second order is used as reference signal for the jitter analysis.
Enables basic measurement "Event Count" and advanced jitter components:
TJ@BER, RJ, TJ, DJ and PJ.
In the "Result" dialog it enables:
"Bathtub" for total jitter @ BER (TJ@BER).
"Step response" for the analyzed source signal.
"Histogram" for RJ, TJ, DJ, DDJ and PJ.
Enables the advance jitter measurement analysis.
Remote command:
ADVJitter<m>:SIGNal:QUICkmeas on page 2262
Reference Level
Opens a dialog to set the reference levels. For details, see Chapter 8.2.4, "Reference
levels", on page 342.
Gating
Opens a dialog to set up the gate of the waveform for the advanced jitter analysis. The
gate limits the measurement to a user-defined range of the waveform.
Advanced jitter and noise (option R&S RTP- K133/K134)

Table of Contents

Other manuals for Rohde & Schwarz RTP044

Related product manuals