Deembedding (Option R&S RTP-K121/-K122)
R&S
®
RTP
1073User Manual 1337.9952.02 ─ 12
19.4.1.1 Proven probe / cable setup
Access: [ App Cockpit] > "Analysis" > "Deembedding" > "Setup" tab: include compo-
nent "Proven Probe" or "Proven Cable" > "Configure" > "Start wizard"
Depending on the steps that you have done before, you can:
Start the complete characterization procedure.
Start the probe or cable measurement in a normalized setup.
Load measurement results.
Select type
Select the DUT to be characterized: probe or cable
Normalize setup
Startes the wizard for the complete normalization and measurement procedure.
Measure probe / cable
Starts the measurement of the probe or cable directly. If the measurement setup has
been normalized just before, the "Measurement" dialog box opens. Otherwize, you are
asked to open a normalization file, or to start with normalization.
Open measurement setup
Opens a file with previously saved measurement results. The "Response" dialog box
opens, and you can check the response plots, and save or apply the S-parameters to
the deembedding setup.
Proven probe, proven cable