Network operation and remote control
R&S
®
RTP
1117User Manual 1337.9952.02 ─ 12
Table 21-12: Bits in the STATus:QUEStionable:ZVCO register
Bit No. Meaning
0 Overload on Z1V1
1 Overload on Z1V2
2 Overload on Z1V3
3 Overload on Z1V4
4 Overload on Z1I1
5 Overload on Z1I2
6 Overload on Z1I3
7 Overload on Z1I4
8 Overload on Z2V1
9 Overload on Z2V2
10 Overload on Z2V3
11 Overload on Z2V4
12 Overload on Z2I1
13 Overload on Z2I2
14 Overload on Z2I3
15 Overload on Z2I4
21.5.4.3 Reset values of the status reporting system
The following table contains the different commands and events causing the status
reporting system to be reset. None of the commands, except *RST and
SYSTem:PRESet, influence the functional instrument settings. In particular, DCL does
not change the instrument settings.
Table 21-13: Resets of the status reporting system
Event Switching on supply
voltage
Power-On-Status-
Clear
DCL,
SDC
(Device
Clear,
Selected
Device
Clear)
*RST or
SYS-
Tem:PRE
Set
STA-
Tus:PRE-
Set
*CLS
Effect 0 1
Clear STB, ESR - yes - - - yes
Clear SRE, ESE - yes - - - -
Clear PPE - yes - - - -
Clear EVENt parts of the regis-
ters
- yes - - - yes
Remote control