Remote control commands
R&S
®
RTP
1174User Manual 1337.9952.02 ─ 12
Example:
:DISPlay:SIGNal:LABel:POSMode C2W1, REL
DISPlay:SIGNal:LABel:HORizontal:RELative:POSition 'Label1', C2W1, 30
DISPlay:SIGNal:LABel:VERTical:RELative:POSition 'Label1', C2W1, 70
Move the label to new relative position: horizontal at 30 % and
vertical at 70 % of the screen.
22.7.2.7 Dialog appearance
DISPlay:EXTended:PORDialogs <PopOutResDialogs>
If an external monitor is connected to the instrument, enables the display of result
boxes on the external monitor.
Parameters:
<PopOutResDialogs> ON | OFF
*RST: OFF
DISPlay:EXTended:POSDialogs
<PopOutSettDialogs>
If an external monitor is connected to the instrument, enables the display of dialog
boxes on the external monitor.
Parameters:
<PopOutSettDialogs> ON | OFF
*RST: OFF
22.7.2.8 Cursor appearance
CURSor<m>:STYLe <Style>
Defines how the cursor is displayed in the diagram.
Suffix:
<m>
.
1..4
Selects the cursor set.
Parameters:
<Style> LINes | LRHombus | VLRHombus | RHOMbus
LINes
The cursors are displayed as lines.
LRHombus
The cursors are displayed as lines. The intersections of the cur-
sors with the waveforms are displayed by rhombus-shaped
points.
VLRHombus
The cursors are displayed only as vertical lines. The intersec-
tions of the cursors with the waveforms are displayed by rhom-
bus-shaped points.
Instrument setup