Remote control commands
R&S
®
RTP
1189User Manual 1337.9952.02 ─ 12
TIMebase:RANGe <AcquisitionTime>
Defines the time of one acquisition, that is the time across the 10 divisions of the dia-
gram: TimeScale*10.
Parameters:
<AcquisitionTime> Range: 250E-12 to 100E+3 (RTO, RTP) | 50E+3 (RTE)
Increment: 1E-12
*RST: 0.5
Default unit: s
Usage: Asynchronous command
TIMebase:DIVisions?
Queries the number of horizontal divisions on the screen. The number cannot be
changed.
Return values:
<HorizDivCnt> Range: 4 to 20
Increment: 2
*RST: 10
Usage: Query only
Asynchronous command
TIMebase:HORizontal:POSition <RescalCenterTime>
Defines the time distance between the reference point and the trigger point (the zero
point of the diagram). The reference point marks the rescaling center of the time scale.
Parameters:
<RescalCenterTime> Range: -100E+24 to 100E+24
Increment: 1E-12
*RST: 0
Default unit: s
Usage: Asynchronous command
TIMebase:REFerence <RescaleCtrPos>
Sets the position of the reference point in % of the screen. The reference point marks
the rescaling center of the time scale. If you modify the time scale, the reference point
remains fixed on the screen, and the scale is stretched or compresses to both sides of
the reference point.
Parameters:
<RescaleCtrPos> Range: 0 to 100
Increment: 1
*RST: 50
Default unit: %
Usage: Asynchronous command