Remote control commands
R&S
®
RTP
1214User Manual 1337.9952.02 ─ 12
TRPRobe:SETup:ATTenuation:MODE <AttenuationMode>
PROBe<m>:SETup:ATTenuation:MODE <PrbAttMd>
Set the mode to MANual if the instrument does not detect the probe.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<PrbAttMd> AUTO | MANual
*RST: AUTO
Usage: Asynchronous command
TRPRobe:SETup:ATTenuation:UNIT
<AttenuationUnit>
PROBe<m>:SETup:ATTenuation:UNIT <AttenuationUnit>
Sets the unit for the connected probe type if PROBe<m>:SETup:ATTenuation:MODE
on page 1214 is set to MANual.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<AttenuationUnit> V | A | W
Voltage probe (V), current probe (A), power probe (W)
*RST: V
Usage: Asynchronous command
TRPRobe:SETup:ATTenuation:MANual <PrbAttMdManual>
PROBe<m>:SETup:ATTenuation:MANual <ManualAttenuation>
Sets the attenuation for the connected probe if PROBe<m>:SETup:ATTenuation:
MODE on page 1214 is set to MANual.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<ManualAttenuation> Range: 100E-6 to 10000
Increment: 0.1
*RST: 1
Default unit: Depends on the selected unit
Usage: Asynchronous command