Remote control commands
R&S
®
RTP
1220User Manual 1337.9952.02 ─ 12
Parameters:
<POffset> In "P" probe mode (PMODe), this offset is used as channel offset
and considered automatically for correction.
Range: -100E+24 to 100E+24
Increment: 1E-3
*RST: 0
Default unit: V
Usage: Asynchronous command
PROBe<m>:SETup:TERM:STATe
<VoltageState>
Activates the instrument control of the termination voltage.
Suffix:
<m>
.
1..4
Parameters:
<VoltageState> ON | OFF
*RST: OFF
Usage: Asynchronous command
PROBe<m>:SETup:TERM:MODE <Mode>
Setting the termination voltage is relevant if you use the R&S RT-ZMA40 SMA module.
You can set a termination voltage to correct the internal 50 termination of the SMA
module by the common mode voltage. To control the termination voltage by the instru-
ment, connect the VT terminal of the R&S RT-ZM probe amplifier to the VT terminal of
the SMA module using the red DC lead (see R&S RT-ZM User Manual).
Suffix:
<m>
.
1..4
Parameters:
<Mode> AUTO | MANual
AUTO
The instrument uses the measured common mode voltage to
control the termination.
MANual
Enter the voltage to be used for termination.
*RST: AUTO
Usage: Asynchronous command
PROBe<m>:SETup:TERM:MEASure?
Returns the measured common mode voltage.
Suffix:
<m>
.
1..4