Remote control commands
R&S
®
RTP
1224User Manual 1337.9952.02 ─ 12
PROBe<m>:DEEMbedding:REMProbeload <RemoveProbeLoad>
Removes the loading of the probe if DEEMbedding<m>:COMPonent<n>:MODE is set
to PROBing.
If the probe loading is removed, you see the signal that would be at the measurement
point if the probe's input impedance is ideal (infinite impedance).
Without removing the loading, you see the real signal at the measurement point,
including the probe loading.
Suffix:
<m>
.
1..4
Selects the channel to which the probe is connected.
Parameters:
<RemoveProbeLoad>ON | OFF
*RST: ON
Usage: Asynchronous command
PROBe<m>:DEEMbedding:LOAD:IMPedance? <SourceImpedance>
Shows the source impedance of the DUT. If an R&S RT-ZM probe is connected, the
value depends on the selected probe mode: common mode, differential, or single-
ended measurement.
Suffix:
<m>
.
1..4
Selects the channel to which the probe is connected.
Return values:
<SourceImpedance> Range: 100E-15 to 1E+9
Increment: 1
*RST: 50
Default unit:
Ω
Usage: Query only
Asynchronous command
22.8.7.7 Predefined probes
TRPRobe:SETup:ATTenuation:DEFProbe...................................................................... 1224
PROBe<m>:SETup:ATTenuation:DEFProbe................................................................... 1224
PROBe<m>:SETup:OFFSet:TOMean............................................................................ 1225
TRPRobe:SETup:ATTenuation:DEFProbe <SelcPredefPrb>
PROBe<m>:SETup:ATTenuation:DEFProbe <PredefinedProbe>
Selects a predefined probe. These are probes that are not recognized automatically
but the parameters of the probe are known to the instrument.
Acquisition and setup