Remote control commands
R&S
®
RTP
1326User Manual 1337.9952.02 ─ 12
Parameters:
<RelativeSpan> Range: 1E-15 to 100
Increment: 0.1
*RST: 1
Default unit: %
Parameters for setting and query:
<DiagramName> Name of the diagram on which the zoom area is based.
<ZoomName> Name of the zoom diagram
Usage: Asynchronous command
LAYout:ZOOM:VERTical:RELative:STARt
<DiagramName>,
<ZoomName>,<RelativeStart>
LAYout:ZOOM:VERTical:RELative:STARt? <DiagramName>, <ZoomName>
Defines the lower limit of the zoom area on the y-axis.
Parameters:
<RelativeStart> Range: 0 to 100
Increment: 0.1
*RST: 0
Default unit: %
Parameters for setting and query:
<DiagramName> Name of the diagram on which the zoom area is based.
<ZoomName> Name of the zoom diagram
Usage: Asynchronous command
LAYout:ZOOM:VERTical:RELative:STOP <DiagramName>,
<ZoomName>,<RelativeStop>
LAYout:ZOOM:VERTical:RELative:STOP? <DiagramName>, <ZoomName>
Defines the upper limit of the zoom area on the y-axis.
Parameters:
<RelativeStop> Range: 0 to 100
Increment: 0.1
*RST: 100
Default unit: %
Parameters for setting and query:
<DiagramName> Name of the diagram on which the zoom area is based.
<ZoomName> Name of the zoom diagram
Usage: Asynchronous command
LAYout:ZOOM:REMove <DiagramName>, <ZoomName>
Removes the specified zoom diagram.