Remote control commands
R&S
®
RTP
1405User Manual 1337.9952.02 ─ 12
<MeasType>
parameter value
(remote control)
Measurement
(manual opera-
tion)
Description/Result Suffix
<n>
UPEakvalue Upper peak value Signal value at the maximum sample value in the
upper half of the histogram
6
LPEakvalue Lower peak value Signal value at the maximum sample value in the
lower half of the histogram
7
HMAXimum Maximum Highest signal value with a probability > 0 8
HMINimum Minimum Lowest signal value with a probability > 0 9
MEDian Median Signal value for which half the samples lie above,
the other half below in the histogram
10
MAXMin Max - Min Range of signal values with a probability > 0 11
HMEan Mean Weighted arithmetic average of the histogram 12
HSTDdev σ (S-dev) Standard deviation of the sample numbers 13
M1STddev Mean ±σ Range between (mean value + standard deviation)
and (mean value - standard deviation)
14
M2STddev Mean ±2*σ Range between (mean value + 3 * standard devia-
tion) and (mean value - 2 * standard deviation)
15
M3STddev Mean ±3*σ Range between (mean value + 3 * standard devia-
tion) and (mean value - 2 * standard deviation)
16
MKPositive Marker + Probabil-
ity %
Marker value (according to the selected probability
domain marker type) plus the defined limit.
Note that the value is restricted to the histogram
range.
17
MKNegative Marker - Probability
%
Marker value (according to the selected probability
domain marker type) minus the defined limit.
Note that the value is restricted to the histogram
range.
18
MEASurement<m>:HISTogram:SELect.......................................................................... 1405
MEASurement<m>:HISTogram:PROBability:TYPE..........................................................1406
MEASurement<m>:HISTogram:PROBability:LIMit........................................................... 1406
MEASurement<m>:HISTogram:LCHeck<n>:VALid.......................................................... 1407
MEASurement<m>:HISTogram:LCHeck<n>:LOWer:LIMit.................................................1407
MEASurement<m>:HISTogram:LCHeck<n>:UPPer:LIMit................................................. 1407
MEASurement<m>:HISTogram:LCHeck<n>:LOWer:MARGin........................................... 1408
MEASurement<m>:HISTogram:LCHeck<n>:UPPer:MARGin............................................ 1408
MEASurement<m>:HISTogram:SELect <HistgRef>
Selects the histogram on which the measurement is based.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
Automatic measurements