Remote control commands
R&S
®
RTP
1452User Manual 1337.9952.02 ─ 12
CALCulate:MATH<m>:FFT:GATE[:STATe] <State>
Enables FFT gating.
Suffix:
<m>
.
1..8
math waveform
Parameters:
<State> ON | OFF
*RST: OFF
Usage: Asynchronous command
CALCulate:MATH<m>:FFT:MAGNitude:LEVel <VerticalMax>
Defines the reference level for dB scaling.
Suffix:
<m>
.
1..8
Math waveform
Parameters:
<VerticalMax> Range: -1E+15 to 1E+15
Increment: 0.01
*RST: 0
Default unit: div
Usage: Asynchronous command
CALCulate:MATH<m>:FFT:MAGNitude:RANGe <Range>
Defines the vertical value range in spectrum mode.
Suffix:
<m>
.
1..8
math waveform
Parameters:
<Range> Range: 1 to 500
Increment: 1
*RST: 100
Default unit: dB
Usage: Asynchronous command
CALCulate:MATH<m>:FFT:MAGNitude:SCALe <MagnitudeScale>
Defines the scaling of the y-axis. The display values are valid for 50Ω termination
impendance.