Remote control commands
R&S
®
RTP
1623User Manual 1337.9952.02 ─ 12
*RST: ANY
TRIGger<m>:CAN:ICONdition <IdOperator>
Sets the operator to set a specific identifier or an identifier range.
Parameters:
<IdOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These condition require one identifier pat-
tern to be set with TRIGger<m>:CAN:IMIN.
INRange | OORange
In range / Out of range: Set the minimum and maximum value of
the range with TRIGger<m>:CAN:IMIN and TRIGger<m>:
CAN:IMAX on page 1623.
*RST: EQUal
TRIGger<m>:CAN:IMIN <IdPattern>
Specifies a message identifier pattern, or sets the the start value of an identifier range.
Parameters:
<IdPattern> Numeric or string pattern, see Chapter 22.4.6, "Bit pattern
parameter", on page 1148.
TRIGger<m>:CAN:IMAX <IdPatternTo>
Sets the the end value of an identifier range if TRIGger<m>:CAN:ICONdition is set
to INRange or OORange.
Parameters:
<IdPatternTo> Numeric or string pattern, see Chapter 22.4.6, "Bit pattern
parameter", on page 1148.
TRIGger<m>:CAN:FDATa:FDF <FDFBit>
Specifies the CAN FD frame format. It corresponds to the EDL bit (extended data
length), which only exists in CAN FD format.
Parameters:
<FDFBit> ONE | ZERO | DC
ONE: CAN FD.
ZERO: CAN.
DC: don't care, the format is not relevant.
*RST: DC
Protocols