EasyManua.ls Logo

Rohde & Schwarz RTP044 - Page 1636

Rohde & Schwarz RTP044
2403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTP
1636User Manual 1337.9952.02 ─ 12
BUS<m>:CAN:FDATa:FRAMe<n>:SCValue?
Returns the stuff bit count modulo 8 value.
Suffix:
<m>
.
1..4
<n> *
Return values:
<StuffCount> Range: 0 to 7
Increment: 1
*RST: 0
Usage: Query only
BUS<m>:CAN:FRAMe<n>:BYTE<o>:STATe?
Returns the state of the specified byte.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
<o> *
Selects the byte number.
Return values:
<State> OK | ERRor | UNDF
UNDF: Undefined
*RST: OK
Usage: Query only
BUS<m>:CAN:FRAMe<n>:BYTE<o>:VALue?
Returns the value of the specified byte.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
<o> *
Selects the byte number.