Remote control commands
R&S
®
RTP
1641User Manual 1337.9952.02 ─ 12
INRange | OORange
In range / Out of range: Set the minimum and maximum value of
the range with SEARch:TRIGger:CAN:DMIN and SEARch:
TRIGger:CAN:DMAX.
*RST: EQUal
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:CAN:DMIN <SearchName>,<DataPattern>
SEARch:TRIGger:CAN:DMIN? <SearchName>
Specifies a data pattern, or sets the the start value of a data pattern range.
Parameters:
<DataPattern> Numeric or string pattern, see Chapter 22.4.6, "Bit pattern
parameter", on page 1148. The string parameter accepts the bit
value X (don't care).
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:CAN:DMAX <SearchName>,<DataPatternTo>
SEARch:TRIGger:CAN:DMAX? <SearchName>
Sets the the end value of an data range if SEARch:TRIGger:CAN:DCONdition is
set to INRange or OORange.
Parameters:
<DataPatternTo> Numeric or string pattern, see Chapter 22.4.6, "Bit pattern
parameter", on page 1148. The string parameter accepts the bit
value X (don't care).
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:CAN:DLCCondition <SearchName>,<DLCOperator>
SEARch:TRIGger:CAN:DLCCondition? <SearchName>
Operator to set the data length code for search.
Parameters:
<DLCOperator> EQUal | GETHan
*RST: GETHan
Parameters for setting and query:
<SearchName>
Protocols