Remote control commands
R&S
®
RTP
1643User Manual 1337.9952.02 ─ 12
INRange = RANGe
In range: Set the minimum and maximum value of the range with
SEARch:TRIGger:CAN:FDATa:DPOSition and SEARch:
TRIGger:CAN:FDATa:DPTO.
*RST: ANY
Parameters for setting and query:
<SearchName> String with the search name
SEARch:TRIGger:CAN:FDATa:DPOSition <SearchName>,<DataPosition>
SEARch:TRIGger:CAN:FDATa:DPOSition? <SearchName>
Defines the first possible start position of the data pattern.
The setting is available in CAN FD option R&S RTP-K9.
Parameters:
<DataPosition> Range: 1 to 57
Increment: 1
*RST: 1
Parameters for setting and query:
<SearchName> String with the search name
SEARch:TRIGger:CAN:FDATa:DPTO <SearchName>,<DataPositionTo>
SEARch:TRIGger:CAN:FDATa:DPTO? <SearchName>
Defines the last possible start position of the data pattern if the position operator
SEARch:TRIGger:CAN:FDATa:DPOPerator defines a range.
The setting is available in CAN FD option R&S RTP-K9.
Parameters:
<DataPositionTo> Range: 8 to 64
Increment: 1
*RST: 8
Parameters for setting and query:
<SearchName> String with the search name
SEARch:TRIGger:CAN:ACKerror <SearchName>,<AckError>
SEARch:TRIGger:CAN:ACKerror? <SearchName>
Searches for acknowledgement errors. An acknowledgement error occurs when the
transmitter does not receive an acknowledgment - a dominant bit during the Ack Slot.
Parameters:
<AckError> ON | OFF
*RST: ON
Parameters for setting and query:
<SearchName>