Remote control commands
R&S
®
RTP
1700User Manual 1337.9952.02 ─ 12
Usage: Query only
BUS<m>:MILStd:WORD<n>:STATus?
Returns the overall state of the selected word.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the word.
Return values:
<FrameState> OK | SYNC | MANC | PAR | GAP | RT
OK: the word is valid.
SYNC: synchronization error occured.
MANC: manchester coding error occured.
PAR: parity error occured.
GAP: timing gap error occured.
RT: remote terminal error occured.
*RST: OK
Usage: Query only
BUS<m>:MILStd:WORD<n>:STOP?
Return the stop time of the selected word.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the word.
Return values:
<FrameStop> Range: -100E+24 to 100E+24
Increment: 100E-12
*RST: 0
Default unit: s
Usage: Query only
BUS<m>:MILStd:WORD<n>:SYMBol?
Returns the label name of the word ID.
Suffix:
<m>
.
1..4
Selects the serial bus.
Protocols