Remote control commands
R&S
®
RTP
1714User Manual 1337.9952.02 ─ 12
Parameters:
<ThresholdLow> Range: -12 to 0
Increment: 0.1
*RST: -5
Default unit: V
22.17.9.2 Trigger
The trigger suffix <m> is always 1 and can be omitted. It selects the trigger event: Only
the A-trigger is available for triggering on serial buses.
To trigger on a serial bus, make sure that:
●
TRIGger<m>:SOURce[:SELect] is set to SBUS.
●
The sources of the serial bus are channel signals: use BUS<m>:...:SOURce
commands.
●
Decoding is enabled: BUS<m>[:STATe] is set to ON.
TRIGger<m>:ARINc:TYPE............................................................................................1714
TRIGger<m>:ARINc:DATA:CONDition............................................................................1714
TRIGger<m>:ARINc:DATA:MIN..................................................................................... 1715
TRIGger<m>:ARINc:DATA:MAX.................................................................................... 1715
TRIGger<m>:ARINc:ERRor:CODing..............................................................................1715
TRIGger<m>:ARINc:ERRor:PARity................................................................................1715
TRIGger<m>:ARINc:LABel:CONDition........................................................................... 1716
TRIGger<m>:ARINc:LABel:MIN.....................................................................................1716
TRIGger<m>:ARINc:LABel:MAX....................................................................................1716
TRIGger<m>:ARINc:MINGap:SELect.............................................................................1716
TRIGger<m>:ARINc:MINGap:BITS................................................................................1716
TRIGger<m>:ARINc:MAXGap:SELect............................................................................1717
TRIGger<m>:ARINc:MAXGap:BITS...............................................................................1717
TRIGger<m>:ARINc:SDI...............................................................................................1717
TRIGger<m>:ARINc:SSM............................................................................................. 1717
TRIGger<m>:ARINc:TYPE <Type>
Sets the trigger type for ARINC 429 analysis.
Parameters:
<Type> STARt | STOP | LABel | ERRor
*RST: STARt
TRIGger<m>:ARINc:DATA:CONDition <DataOperator>
Sets the condition for the data. You can define an exact data patterm or a data range.
Parameters:
<DataOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
Protocols