Remote control commands
R&S
®
RTP
1797User Manual 1337.9952.02 ─ 12
Suffix:
<m>
.
1..3
Parameters:
<SrcAddrOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data pattern
to be set with the corresponding TRIGger<m>:TBTO:
SADDress:MIN command.
INRange | OORange
In range/Out of range: set the minimum and maximum value of
the range with TRIGger<m>:TBTO:SADDress:MIN and
TRIGger<m>:TBTO:SADDress:MAX.
*RST:
EQUal
TRIGger<m>:TBTO:SADDress:MAX
<SrcAddrPattMax>
Sets the end value of the source address range, if TRIGger<m>:TBTO:SADDress:
CONDition is set to INRange or OORange.
Suffix:
<m>
.
1..3
Parameters:
<SrcAddrPattMax> Numeric or string pattern, see Chapter 22.4.6, "Bit pattern
parameter", on page 1148. The string parameter accepts the bit
value X (do not care).
TRIGger<m>:TBTO:SADDress:MIN <SrcAddrPattMin>
Specifies the source address bit pattern, or sets the start value of a pattern range.
Suffix:
<m>
.
1..3
Parameters:
<SrcAddrPattMin> Numeric or string pattern, see Chapter 22.4.6, "Bit pattern
parameter", on page 1148. The string parameter accepts the bit
value X (do not care).
22.17.12.4 Search settings
SEARch:TRIGger:TBTO:CRC:CONDition.......................................................................1798
SEARch:TRIGger:TBTO:CRC:MAX............................................................................... 1798
SEARch:TRIGger:TBTO:CRC:MIN................................................................................ 1799
SEARch:TRIGger:TBTO:DADDress:CONDition...............................................................1799
SEARch:TRIGger:TBTO:DADDress:MAX.......................................................................1799
SEARch:TRIGger:TBTO:DADDress:MIN........................................................................ 1800
Protocols