Remote control commands
R&S
®
RTP
1880User Manual 1337.9952.02 ─ 12
Suffix:
<m>
.
1..3
Parameters:
<CondOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one pattern to
be set with the corresponding TRIGger<m>:MPHY:LWONe:MIN
command.
INRange | OORange
In range/Out of range: set the minimum and maximum value of
the range with TRIGger<m>:MPHY:LWONe:MIN and
TRIGger<m>:MPHY:LWONe:MAX.
*RST:
EQUal
TRIGger<m>:MPHY:LWONe:MIN
<LCCWORD1Min>
Specifies the WORD 1 pattern, or sets the start value of a pattern range.
Suffix:
<m>
.
1..3
Parameters:
<LCCWORD1Min> Numeric or string pattern, see Chapter 22.4.6, "Bit pattern
parameter", on page 1148. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:MPHY:LWONe:MAX <LCCWORD1Max>
Sets the end value of the WORD 1 range if TRIGger<m>:MPHY:LWONe:CONDition
is set to INRange or OORange.
Suffix:
<m>
.
1..3
Parameters:
<LCCWORD1Max> Numeric or string pattern, see Chapter 22.4.6, "Bit pattern
parameter", on page 1148. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:MPHY:LWTWo:CONDition <CondOperator>
Sets the condition for WORD 2. You can define an exact value or a value range.