EasyManua.ls Logo

Rohde & Schwarz RTP044 - Page 1906

Rohde & Schwarz RTP044
2403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTP
1906User Manual 1337.9952.02 ─ 12
SEARch:TRIGger:MPHY:TC <SearchName>,<TC>
SEARch:TRIGger:MPHY:TC? <SearchName>
Specifies the TC bit pattern.
Parameters:
<TC> Numeric or string pattern, see Chapter 22.4.6, "Bit pattern
parameter", on page 1148. The string parameter accepts the bit
value X (don't care).
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:MPHY:DATA:DCONdition
<SearchName>,<DataOperator>
SEARch:TRIGger:MPHY:DATA:DCONdition? <SearchName>
Sets the condition for the data. You can define an exact value or a value range.
Parameters:
<DataOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data pattern
to be set with the corresponding SEARch:TRIGger:MPHY:
DATA:DMIN command.
INRange | OORange
In range/Out of range: set the minimum and maximum value of
the range with SEARch:TRIGger:MPHY:DATA:DMIN and
SEARch:TRIGger:MPHY:DATA:DMAX.
*RST:
EQUal
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:MPHY:DATA:DMIN <SearchName>,<DataMin>
SEARch:TRIGger:MPHY:DATA:DMIN? <SearchName>
Specifies the data bit pattern, or sets the the start value of a pattern range.
Parameters:
<DataMin> Numeric or string pattern, see Chapter 22.4.6, "Bit pattern
parameter", on page 1148. The string parameter accepts the bit
value X (don't care).
Parameters for setting and query:
<SearchName>
Protocols

Table of Contents

Other manuals for Rohde & Schwarz RTP044

Related product manuals