Triggers
R&S
®
RTP
216User Manual 1337.9952.02 ─ 12
Source
Selects the source of the trigger signal for the current trigger condition. The trigger
source works even if it is not displayed in a diagram. It must be synchronized to the
signal to be displayed and analyzed.
The trigger source can be:
Channel 1...4: An analog input channel
Extern: External analog signal connected to the external trigger input
The external trigger source is supported for the "Single event" sequence. It is not
available if a longer trigger sequence is selected, or if qualification is enabled.
Line: The instrument generates the trigger from the AC power input and synchroni-
zes the signal to the AC power frequency. Use this source if you want to analyze
signals related to the power line frequency, such as lighting equipment and power
supply devices.
Serial bus, D0...D15, Logic, Parallel bus 1...4:
If options with trigger functionality are installed, the variety of trigger sources is
enhanced with specific trigger sources.
Available sources depend on the trigger sequence setting. If "Single event" is selected,
all inputs (analog input channels, serial and parallel buses, digital channels) can be
used as trigger source. If any other trigger sequence is selected, only channel inputs
Ch1...4 can be set as trigger source, and all other input sources are disabled.
See also: Chapter 6.8, "Sequence", on page 246
Remote command:
TRIGger<m>:SOURce[:SELect] on page 1253
Type
Selects the trigger type specific for each condition in a trigger sequence. The current
trigger type is shown on the button.
The following trigger types are available:
Edge, see page 217
Glitch, see page 218
Width, see page 219
Runt, see page 220
Window, see page 222
Timeout, see page 223
Interval, see page 224
Slew rate, see page 226
Setup & Hold trigger, see page 227
State, see page 229
Pattern, see page 230
High-speed serial pattern trigger (Option R&S RTP-K140/K141), see page 253
Line trigger, see page 237
Restrictions:
If the external trigger input is used as trigger source, the analog edge trigger is the
only available trigger type.
For the R-trigger (reset), the trigger types and settings are restricted dependent on
the A and B trigger settings. The instrument provides only possible, reasonable
combinations.
Trigger types

Table of Contents