EasyManua.ls Logo

Rohde & Schwarz RTP044 - Page 2246

Rohde & Schwarz RTP044
2403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTP
2246User Manual 1337.9952.02 ─ 12
Usage: Event
Asynchronous command
22.20.2.2 Measurements
TDRT:CALibration:USECaldata..................................................................................... 2246
TDRT:BW....................................................................................................................2246
TDRT:DOMain............................................................................................................. 2246
TDRT:PERMittivity........................................................................................................2247
TDRT:SIGNal<m>STATe...............................................................................................2247
TDRT:SIGNal<m>:SCALe.............................................................................................2247
TDRT:SIGNal<m>:OFFSet............................................................................................2248
TDRT:SIGNal<m>:POSition...........................................................................................2248
TDRT:SIGNal<m>:DATA:HEADer?.................................................................................2248
TDRT:SIGNal<m>:DATA:VALues?................................................................................. 2249
TDRT:CALibration:USECaldata <UseCalibrationDat>
Disables or enables the frequency response calibration. Without calibration, the mea-
sured signals are only rescaled, and you can see the impact of the frequency response
calibration on the time domain waveform.
The command is not relevant when measuring with default values.
Parameters:
<UseCalibrationDat> ON | OFF
*RST: ON
Usage: Asynchronous command
TDRT:BW <Bandwidth>
Sets the bandwidth of the DUT.
Parameters:
<Bandwidth> Range: 2E+9 to 16E+9
Increment: 10
*RST: 4E+9
Default unit: Hz
Usage: Asynchronous command
TDRT:DOMain <OutputDomain>
Sets the domain of the x-axis: time or distance (in m or inch). If you select distance,
enter also the permittivity with TDRT:PERMittivity.
Pulse source and TDR/TDT analysis (options R&S
RTP-B7/K130)

Table of Contents

Other manuals for Rohde & Schwarz RTP044

Related product manuals