Remote control commands
R&S
®
RTP
2293User Manual 1337.9952.02 ─ 12
Parameters:
<State> ON | OFF
*RST: OFF
Usage: Asynchronous command
DEEMbedding<m>:ADD <Index>, [<Type>]
Adds a component to the measurement setup.
Suffix:
<m>
.
1..4
Selects the channel that is configured for deembedding.
Setting parameters:
<Index> Defines the place of the component in the measurement path.
The DUT is always the first component (index = 1), the probe or
input is the last component.
<Type> PROBe | FIXTure | CABLe | DUT | ADAPter | INPut | OTHer |
RTZA10 | RTZA16 | RTZA17 | PP | PC | COPRobes
RTZA10 | RTZA16 | RTZA17
SMA adapters and cable offered by Rohde & Schwarz
PP | PC
Proven probe, proven cable
Usage: Setting only
Asynchronous command
DEEMbedding<m>:REMove <Index>
Deletes the component with the indicated index from the measurement setup.
Suffix:
<m>
.
1..4
Selects the channel that is configured for deembedding.
Setting parameters:
<Index> Defines the place of the component in the measurement path.
The DUT is always the first component (index = 1), the probe or
input is the last component.
Usage: Setting only
Asynchronous command
DEEMbedding<m>:CCOunt?
Returns the total number of components in the measurement setup, including disabled
components.
Deembedding (Option R&S