Triggers
R&S
®
RTP
251User Manual 1337.9952.02 ─ 12
2. In the signal bar, select if the signal must violate the zone to cause a trigger (Must
Intersect), or vice versa (Must Not Intersect).
3. Tap the corner points of the zone on the screen.
4. If all points are set, tap "Finish zone" in the signal bar.
"Zone1" is shown in the expression field.
5. If you want to define another zone, select the logic combination of the zones AND
(both zones) or OR (either zone).
6. To create the second zone, repeat steps 3 and 4.
7. To define more zones, repeat steps 5 and 6.
Use existing masks for zone trigger
Prerequisites: At least one mask is created and visible on the display.
1. Tap the "Zone trigger" icon on the toolbar.
2. In the signal bar, select if the signal must violate the zone to cause a trigger (Must
Intersect), or vice versa (Must Not Intersect).
3. Tap the mask on the display.
4. Tap "Use selected mask" in the signal bar.
"MT1" is shown in the expression field.
5. If you want to add another mask test or zone, select the logic combination of the
zones AND (both zones) or OR (either zone).
6. To add another mask, tap the mask on the display.
Tap "Use selected mask" in the signal bar.
7. To add another zone, tap the corner points of the zone on the screen.
If all points are set, tap "Finish zone" in the signal bar.
If a zone trigger is already defined, you can also add a new mask to the zone trigger
when creating the mask.
Intersect
Defines if the signal must intersect the zone to allow the instrument to trigger, or if it
must not intersect the zone.
Combine
Sets the logic combination of two zones.
Use selected mask
Includes the selected zone or mask in the zone expression.
Zone trigger

Table of Contents

Other manuals for Rohde & Schwarz RTP044

Related product manuals