Remote control commands
R&S
®
RTP
1393User Manual 1337.9952.02 ─ 12
MEASurement<m>:SPECtrum:CPOWer:BANDwidth <ChPowBw>
Defines the bandwidth over which the channel power is calculated.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
Parameters:
<ChPowBw> Range: 0 to 4E+9
Increment: 1
*RST: 0
Default unit: Hz
Usage: Asynchronous command
MEASurement<m>:SPECtrum:OBANdwidth <OccupiedBW>
Defines the percentage of the total power used to determine the occupied bandwidth.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
Parameters:
<OccupiedBW> Range: 0.1 to 99.9
Increment: 1
*RST: 20
Default unit: %
Usage: Asynchronous command
MEASurement<m>:SPECtrum:CPOWer:CFRequency <ChPowCtrFreq>
Defines the center frequency from which the channel power is calculated over the
specified bandwidth.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
Parameters:
<ChPowCtrFreq> Range: 0 to 6E+9
Increment: 1
*RST: 0
Default unit: Hz
Usage: Asynchronous command